T.A.D. Therapy Alignment Device
710-040
T.A.D. Therapy Alignment Device.
T.A.D. Therapy Alignment Device.
T.A.D. Therapy Alignment Device.
One simple device can align your Linac, ODI, CT, Lasers and Couch.
Perform the following tests with just one tool:
• Optical Distance Check Over 20 cm Range
• Collimator 90° Rotation Test
• Lasers Test: Side, Overhead and Sagittal
• Couch Height Test Over 20 cm Range
• Couch Vertical Travel Wander
• Couch 90° Left or Right Rotation
• CT Test Cut for Couch Center Alignment and Height
• CT Side Laser Height Check Over 20 cm
• CT Laser Height Test Over 20 cm
• CT Couch Height Vertical Travel
• CT Couch Height, Vertical Travel, CT Cut Stays on Same Spot to Indicate Gantry Vertical
• Gantry 90° Right and Left Test with Radiation and Gantry Vertical Test Using Film or Image Receptor
• Light Field vs. Radiation Field 10 cm x 10 cm and 20 cm x 20 cm with divergent tungsten pins
• CT Distance Alignment Check Over 0 cm, 8 cm, 13 cm and 21 cm Viewing 0.75 mm Aluminum Balls (Inside Balls 5 cm apart and outside Balls 21 cm apart)
Therapy Alignment Device Specifications
Leveling Plate Base
• 8.5" x 8.5" x 3/4" thick (21.6 x 21.6 x 1.9 cm) black acrylic with a 1/16" (0.16 cm) white engraving material top
• Three leveling screws with rubber tips to prevent movement
• One bubble level attached to top with screws
• Two half-moon side plates, scribed with black dashes that align to top of leveling plate and center markings on leveling plate
Scribing on Leveling Plate
• Central axis to 20 cm
• 10 cm x 10 cm field
• 20 cm x 20 cm field
Accelerator Markers in Leveling Plate
• Eight 1.6 mm diameter x 12.5 mm tungsten pins mounted divergently at the four corners of the 10 cm and 20 cm fields
• One removable tungsten pin in the center
• Two 1.6 mm diameter tungsten pins are horizontally half-way embedded in the scribe lines 3 mm from the side plates
CT Markers in Leveling Plate
• Four 1/32" (0.08 cm) diameter aluminum balls imbedded flush with surface at 2.5 cm and 10.5 cm from central axis in one plane
Note: Tungsten pin must not be in the center hole for CT
ODI Scale Specifications
Base
• 2" x 2" x 0.875" thick (5.08 x 5.08 x 2.22 cm) black acrylic base
• 2" W x 8.125" H x 1/8" thick (5.08 x 20.64 x 0.32 cm) white/black/white engraving material attached to base
Scribing on ODI Scale
• Vertical dashed line
• Horizontal lines at 5.0 cm, 10.0 cm. 15.0 cm and 20.0 cm from bottom of scale