CT Chamber PTW 30017
300-565
CT Chamber PTW 30017
Note: This PTW CT chamber is designed to fit into PTW CTDI Phantoms specifically, and may not fit perfectly into third-party CTDI phantoms.
CT Chamber PTW 30017
Note: This PTW CT chamber is designed to fit into PTW CTDI Phantoms specifically, and may not fit perfectly into third-party CTDI phantoms.
CT Chamber PTW 30017.
The CT chamber is a vented cylinder chamber designed for dose length product and dose length product rate measurements in computed tomography according to the amendment to IEC 60601-2-44.
• Pencil type chamber for measurements free in air
• Provides a sensitive measuring length of 30 cm
• Shows a homogeneous response over the whole chamber length
Specifications
Type of Product: vented pencil type chamber
Application: dosimetry in computed tomography
Measuring Quantity: air kerma length product, exposure length product
Reference Radiation Quality: 120 kV, HVL 8.4 mm Al (RQT9)
Nominal Sensitive Volume: 9.3 cm3
Design: not waterproof, vented, pencil type
Reference Point: chamber center
Direction of Incidence: radial
Nominal Response: 13 nC/(Gy ··cm)
Chamber Voltage: -100 V nominal, ± 500 V maximal high voltage to be connected only with active current-limiting device (Imax< 0.5 mA)
Energy Response: < ± 5 % for 50 - 150 kV
Leakage Current: < ± 10 fA
Cable Leakage: < 1pC/(Gy ··cm)
Materials and Measures
Wall Material: 1 mm PMMA, graphite coated
Wall Area Density: 119 mg/cm2
Dimension of Sensitive Volume: radius 3.5 mm, length 300 mm
Electrode: Al tube, graphite outer diameter 3 mm
Ion Collection Efficiency at Nominal Voltage:
Ion collection time: 274 µs
Max. dose rate for > 95.0 % saturation: 12.4 Gy/s
Max. dose per pulse for > 95.0 % saturation: 2.26 mGy
Useful ranges
Chamber Voltage: ± 100 - 400 V
Radiation Quality: 50 - 150 kV electrons
Temperature: 10° to 40° C, 50° to 104°F
Humidity: 10 to 80 %, max 20 g/m3
Air pressure: 700 - 1060 hPa
Note: This PTW CT chamber is designed to fit into PTW CTDI Phantoms specifically, and may not fit perfectly into third-party CTDI phantoms.
Please specify which connection end is needed